0.635cm Radius (TN-RD-56-0.63) Full Photon Range & 15MeV - 18MeV Electrons MOSFET dosimeters have only 0.8mm inherent build-up, which allows for flexibility in measuring a surface dose as well as the dose at Dmax. In order to measure a dose at Dmax, build-up is required. Measuring accurate doses at depth is of primary concern for the physics community when performing patient dosimetry. Precise dose measurement verifies that the target volume area has been treated with the appropriate dose, while sparing surrounding tissue and organs. Current tissue-equivalent build-up caps require a different size for each energy. Best Medical Canada offers a custom designed Build-up Cap for use with MOSFET Dosimeters, which allows for depth dose measurements over a wide range of energies.
Why Brass? Brass is a metal alloy containing mainly copper and zinc compounds. Due to its high density (8.5g/cm³) and to its high Z number (Z ~ 30), it provides the minimal amount of metal needed to achieve full build-up at Dmax for a range of photon energies (4, 6, 10 and 18MV) and some electron energies (15-18MeV)* at a very practical size. *Note: For low electron energies, it is suggested that no build-up is used. However, if desired, one can use the 1.5cm radius tissue equivalent build-up cap (model TN-RD-55-1.5). Click here to view the Brass Build-up Cap Datasheet (PDF) Tissue Equivalent Manufactured exclusively for Best Medical Canada by CIRS using their Plastic Water™,
these build-up caps come in sizes of 1cm, 1.5cm, and 2cm (models TN-RD-55-1, TN-RD-55-1.5,
TN-RD-55-2 respectively) for various beam energies. The build-up caps have grooves
specifically made for MOSFET dosimeters, allowing accurate measurements.
|
|
|||||||||||||||||||||
|
© 2008-2012
Best Medical Canada Ltd., a member of
TeamBest™ |
|
|||||||||||||||||||||